UP↑SCALE › LABS, QUALITY & METROLOGY › SEALED RECORDS & CHECKS › CALCULATION VALIDATION (IQ/OQ)
Validation of the calculations (IQ/OQ) with reference cases
Labs and service companies have to show that their software calculates correctly. The ENTERPRISE & COMPLIANCE edition runs reference cases worked out by hand from the standards - computed by the UP↑SCALE firmware - and makes a sealed validation record.
- Reference cases worked out by hand from the standards
- Computed by the UP↑SCALE firmware itself
- A record with code and firmware fingerprints
- Sealed raw data for your quality system
In short
UP↑SCALE is one scale app for the scale or balance you already have. Its ENTERPRISE & COMPLIANCE edition validates its calculations: RUN VALIDATION computes reference cases worked out by hand from OIML R 76-1, USP <41>, ISO 8655 and more on the connected UP↑SCALE and makes a sealed validation record with the fingerprints of the code. It is no certificate. From 13.00 EUR a month.
How it works
Laboratories and service companies have to show that the software they use calculates correctly. The ENTERPRISE & COMPLIANCE edition has a box VALIDATION OF THE CALCULATIONS (IQ/OQ) at the end of the ☑ TEST page:
- RUN VALIDATION computes the reference cases below with the code that judges your tests and compares every result with the value worked out by hand from the standard. The UP↑SCALE runs the tests itself, so the cases of the tests (classes, mpe, loads, errors, repeatability, eccentricity, calibration check, creep, zero-setting, stability, discrimination, test weights, linearity, routine check), of the density mode, of the fill weighing (the ℮ / FPackV and HB 133 limits, the verdict, the state of a fill), of the uniformity of mass, of the pipettes (water and air density, Z factor, volumes, uncertainty budget, decision rule), of the coffee modes, of the amount of a sale, of the lot check of ℮ packs, of the US package check (NIST HB 133), of the statistics page (its numbers and the masses of the combination weighing), of the pipette and glassware tables (the limits MPE-, the balance BAL-, the test volumes, the deviations from the standard DEV-, the glassware GLS-, the CMC CMC-, the U of an instrument in the budget) and of the air buoyancy correction (AB-) are sent to the connected UP↑SCALE one by one and computed by the UP↑SCALE firmware; the other cases (the lab's own limits table as text, the makers' specifications of the pipette library, the instrument of an imported row, the readings of other makers' sensors, the reading of the line names of a combination weighing) are computed by the app. Connect your UP↑SCALE first; the run takes a few seconds.
- VALIDATION RECORD (PRINT / PDF) and DOWNLOAD RECORD (HTML) make a record for your validation file: date, app version, a code fingerprint of the app (SHA-256 of its tested functions and the cases) and a firmware fingerprint (SHA-256 of the UP↑SCALE firmware that computed the cases of the tests) - another app version or another firmware gives another fingerprint, so you can show later which code was validated - browser and device, the scale (name, firmware, Device ID, edition), all cases with expected and calculated result, and lines for performed / reviewed / approved by. Like all reports of the edition it gets your company header, a report number of the scale and an integrity code.
- EXPORT DATA (CSV / JSON) saves the same run as raw data for a quality management system: the IQ data (date, app version, code and firmware fingerprint, browser, scale) and one row per case with expected and calculated result and pass/fail. The file is sealed by the scale (CSV with its
.seal.json), so it can be checked above. With the LIMS forwarding the app can also send every run as JSON to the lab's own server. - What it does not cover: the scale itself - the reading of the load cell, the filters, the rounding, the stability detection and the tare. These are tested on the scale with traceable test weights, using the test suite (repeatability, accuracy & linearity, eccentricity, tare and what your procedure asks for); attach its FINAL REPORT to the validation record. Run the validation again after every app or firmware update (a fingerprint changes) and on every device you use for tests.
- It is no certificate: it shows that the UP↑SCALE firmware and the app compute the cases as expected on your devices. The cases are worked out from OIML R 76-1:2006 (the routine check also from USP <41> and Ph. Eur. 2.1.7, the air buoyancy correction from OIML D 28 and OIML R 111-1, the uniformity of mass from Ph. Eur. 2.9.5, the pipettes from ISO 8655-6:2022, ISO/TR 20461:2023, EURAMET cg-19, NIST SOP 14 and ILAC P14) by us; you can check every one of them with the working below. UP↑SCALE is not affiliated with OIML, USP, the EDQM, ISO, ILAC, EURAMET, NIST or any authority.
Where the reference cases come from
The calculations: the reference cases below are worked out by hand from OIML R 76-1:2006 (the cases of the routine check, RT-, from USP <41> and Ph. Eur. 2.1.7; a scale without an accuracy class, U-, against the maker's tolerance from its data sheet (± g + % of the load, no limits of instruments in use, the limits of the standard itself not judged), worked out by hand; the air buoyancy correction, AB-, from OIML D 28 and OIML R 111-1, computed by the UP↑SCALE firmware; the uniformity of mass of the compounding assistant, UM-, from Ph. Eur. 2.9.5; the calibration check, CAL-, from OIML R 76-1 - its quick verdict uses the 0.10 % accuracy limit of USP <41> for acceptable, half of it for OK and our own minimums of 2 d and 5 d; the amount of a sale, PR-, price x weight as the scale gets them, commercial rounding to the cent after DIN 1333, computed by the UP↑SCALE firmware; the lot check of ℮ packs, LOT-, the sample plan and the verdict after 76/211/EEC annex II (the double plan with its 2nd sample, the destructive plan, the 100 % check of a batch under 100, the mean criterion with its factors 0.503 / 0.379 / 0.640 on its own sample, the packs below T1 and T2, an incomplete sample) and after OIML R 87:2016 (the sample plan of Table 2 / Annex I with the SCF, the total inspection of a small lot, the tolerable deficiency of Table 1 with its rounding), computed by the UP↑SCALE firmware; the US package check, HB-, from NIST Handbook 133-2026, Category A - packs labeled by weight, count and volume, with the moisture allowances and the tare sample, computed by the UP↑SCALE firmware; the pipettes, PIP-, from ISO 8655-6:2022 with the water density of Tanaka et al. (Metrologia 38, 2001) for air-free and for air-saturated water (EURAMET cg-19 eq. 3), the air density of OIML R 111-1 (the simplified CIPM formula of EURAMET cg-19) and the uncertainty along ISO/TR 20461:2023 with the handling of the instrument of EURAMET cg-19 (6.3.7.2) and the reference temperature 20 / 27 °C, the volume model also against the two published examples of EURAMET Calibration Guide No. 19 v4.1 (2025) and the model of NIST SOP 14 (NIST IR 7383, 2019); the pipette limits, MPE-, from public secondary sources of ISO 8655-2:2022, with the lab's own table (multichannel columns, limits at a test volume), and for piston burettes from ISO 8655-3:2022 table 1 and formula 1 (the official sample of the standard) and for manual burettes and bottle-top dispensers from maker procedures of ISO 8655-3 / ISO 8655-5; the makers' specifications of the pipette library, LIB-, from the makers' operating manuals as printed (the limits a pipette of the model gets at a test volume, the test volumes of a calibration, a maker printing % only, a multichannel pipette, a volume between two rows); the balance table, BAL-, and the deviations from the standard, DEV-, from ISO 8655-6:2022 itself (5.2, 5.5, 6, 7.2, 8.1.1, 8.1.2 and clause 4); the volumetric glassware, GLS-, the class tolerances of ISO 1042, 648, 835, 385 and 4788 from a maker's catalogue and those of ASTM E288, E969, E1293, E287 and E1272 / USP <31> from a maker's catalogue and USP, the balance and the room rules from the ISO/DIS 4787 sample (ISO 4787:2021), and in the pipette cases PIP-31 to PIP-36 the setting of the meniscus of EURAMET cg-19 (6.3.7.1) and the expansion of borosilicate and soda-lime glass; the CMC, CMC-, after ILAC P14 (the pipette limits, the balance table, the test volumes, the deviations, the glassware tables and the CMC are computed by the UP↑SCALE firmware; the lab's own table as text, the makers' specifications and the instrument of an imported row by the app); the recipes, solutions and the compounding assistant, SOL- (the goals from the shares, the recalculation from one row, the state of a row with its tolerance, 1 d and the air buoyancy correction, and the band the scale accepts - a tolerance narrower than d is flagged), computed by the UP↑SCALE firmware; the coffee modes, COF- (the flow rate of the scale, the yield, the targets and the result of a pour-over, the coffee in the jug at a lift-off, the water of the pours and steps of a recipe, the step at a time, the coffee to expect in the jug, how the scale learns the water the coffee keeps, the dial-in hint of a shot and the extraction yield with its place in the classic box of the brewing control chart, the pace of a timed pour, the countdown to stop pouring and the score of the pour trainer, the time from the START to the first drops and the peak flow of a shot, the dose in a remembered portafilter or dosing cup), computed by the UP↑SCALE firmware; the sensors, SEN-, how the app reads the values of thermometers, hygrometers and barometers of other makers from the published formats (the Bluetooth SIG GATT Specification Supplement for Environmental Sensing and Health Thermometer, the RuuviTag data format 5 with the test vectors of the maker, the BTHome v2 format, the firmware documentation of the Nordic Thingy:52, open-source decoders for the Xiaomi, Aranet, pvvx and Govee formats; the Modbus over serial line specification for the requests, their CRC and the register values; the MQTT 3.1.1 specification for the length bytes and topic filters) and the U of an instrument in the pipette budget (the bigger of the lab's standard uncertainty and U / 2 of the instrument); the units, UNIT- and WU-, from the exact definitions of NIST SP 811 / NIST Handbook 44 appendix C (lb 453.59237 g, oz 28.349523125 g, troy ounce 31.1034768 g, grain 64.79891 mg, tola = 180 grains, pennyweight = 24 grains, metric carat 0.2 g); the statistics page, ST-, with the NIST Statistical Reference Datasets (NumAcc1) and cases worked out by hand, computed by the UP↑SCALE firmware; the combination weighing of the statistics page, WD- (the masses of objects weighed alone and together by least squares, their standard uncertainty and the reading of the line names), with the NIST Statistical Reference Datasets for linear regression without a constant (NoInt1, NoInt2) and cases worked out with exact fractions - the masses computed by the UP↑SCALE firmware, the line names read by the app). The lab's validation record shows the code fingerprint of the app version it used; a different app version has a different fingerprint.
The reference cases
The reference cases (loads in g, errors and limits in d = the scale interval; "checks" lists each check of a test as [error, limit or uncertainty]; status: meets / maybe = cannot be decided / inuse = within the limit for instruments in use / fails; u = uncertainty, level: ok / maybe / warn = only within the limit for instruments in use / bad; verdict of the calibration check: ok / acceptable / recalibrate):
| Case | Clause | Input | Expected | Working |
|---|---|---|---|---|
| T3-01 | R 76-1 Table 3 | classes {"d":0.1,"e":0.1,"max":3000} | ["II"] | n = 3000 / 0.1 = 30000. I needs n >= 50000: no. II with e >= 0.1 g needs 5000 <= n <= 100000: yes. III with 0.1 <= e <= 2 g needs n <= 10000: no. IIII needs e >= 5 g: no. |
| T3-02 | R 76-1 Table 3 | classes {"d":0.01,"e":0.01,"max":300} | ["II"] | n = 300 / 0.01 = 30000. I: n < 50000. II with 0.001 <= e <= 0.05 g needs 100 <= n <= 100000: yes. III needs e >= 0.1 g: no. |
| T3-03 | R 76-1 Table 3 | classes {"d":1,"e":1,"max":3000} | ["III"] | n = 3000. II with e >= 0.1 g needs n >= 5000: no. III with 0.1 <= e <= 2 g needs 100 <= n <= 10000: yes. |
| T3-04 | R 76-1 Table 3 | classes {"d":5,"e":5,"max":3000} | ["III","IIII"] | n = 600. III with e >= 5 g needs 500 <= n <= 10000: yes. IIII with e >= 5 g needs 100 <= n <= 1000: yes. |
| T3-05 | R 76-1 Table 3 | ref {"d":5,"e":5,"max":3000} | "IIII" | Classes III and IIII fit (T3-04); without a class typed in, the loosest class that fits (IIII) is the reference. |
| T3-06 | R 76-1 Table 3 | classes {"d":5,"e":5,"max":2000} | ["IIII"] | n = 400. III with e >= 5 g needs n >= 500: no. IIII needs 100 <= n <= 1000: yes. |
| T3-07 | R 76-1 Table 3 | classes {"d":0.001,"e":0.001,"max":100} | ["I","II"] | n = 100000. I needs e >= 0.001 g and n >= 50000: yes. II with 0.001 <= e <= 0.05 g needs n <= 100000: yes (limit included). |
| T3-08 | R 76-1 Table 3 | classes {"d":0.1,"e":0.1,"max":200} | ["III"] | n = 2000. II with e >= 0.1 g needs n >= 5000: no. III with 0.1 <= e <= 2 g needs 100 <= n <= 10000: yes. |
| T3-09 | R 76-1 Table 3 | classes {"d":1,"e":1,"max":10000} | ["II","III"] | n = 10000. II with e >= 0.1 g: 5000 <= 10000 <= 100000: yes. III: n <= 10000: yes (limit included). |
| T3-16 | R 76-1 Table 3 | classes {"d":0.1,"e":0.1,"max":8000} | ["I","II"] | n = 80000. I needs e >= 0.001 g and n >= 50000: yes. II with e >= 0.1 g needs 5000 <= n <= 100000: yes. III: n > 10000. |
| T3-10 | R 76-1 Table 3, 3.4.3 | min {"d":0.001,"e":0.001,"max":100,"cls":"I"} | 0.1 | Class I: Min = 100 e, with e replaced by d (3.4.3): 100 x 0.001 g = 0.1 g. |
| T3-11 | R 76-1 Table 3, 3.4.3 | min {"d":0.01,"e":0.01,"max":300,"cls":"II"} | 0.2 | Class II with 0.001 <= e <= 0.05 g: Min = 20 d = 20 x 0.01 g = 0.2 g. |
| T3-12 | R 76-1 Table 3, 3.4.3 | min {"d":0.1,"e":0.1,"max":3000,"cls":"II"} | 5 | Class II with e >= 0.1 g: Min = 50 d = 50 x 0.1 g = 5 g. |
| T3-13 | R 76-1 Table 3, 3.4.3 | min {"d":1,"e":1,"max":3000,"cls":"III"} | 20 | Class III: Min = 20 d = 20 g. |
| T3-14 | R 76-1 Table 3, 3.4.3 | min {"d":5,"e":5,"max":3000,"cls":"IIII"} | 50 | Class IIII: Min = 10 d = 50 g. |
| T3-15 | R 76-1 Table 3, 3.4.3 | min {"d":0.1,"e":1,"max":3000,"cls":"III"} | 2 | e = 1 g, d = 0.1 g (an RS-232 adapter with e typed in). Class III: Min = 20 e, but 3.4.3 replaces e by d: 20 x 0.1 g = 2 g. |
| T6-01 | R 76-1 Table 6 | mpe {"d":0.1,"e":0.1,"max":3000,"cls":"II","load":500} | 0.5 | m = 500 / 0.1 = 5000 e. Class II: 0 <= m <= 5000 -> 0.5 e = 0.5 d. |
| T6-02 | R 76-1 Table 6 | mpe {"d":0.1,"e":0.1,"max":3000,"cls":"II","load":500.1} | 1 | m = 5001 e. Class II: 5000 < m <= 20000 -> 1 e = 1 d. |
| T6-03 | R 76-1 Table 6 | mpe {"d":0.1,"e":0.1,"max":3000,"cls":"II","load":2000} | 1 | m = 20000 e. Class II: 5000 < m <= 20000 -> 1 e (limit included). |
| T6-04 | R 76-1 Table 6 | mpe {"d":0.1,"e":0.1,"max":3000,"cls":"II","load":2000.1} | 1.5 | m = 20001 e. Class II: 20000 < m <= 100000 -> 1.5 e. |
| T6-05 | R 76-1 Table 6 | mpe {"d":0.1,"e":0.1,"max":3000,"cls":"II","load":2000.0000000003} | 1 | A load of 2000 g that came out of a unit conversion (e.g. typed in oz) with a rounding of 3e-10 g is still m = 20000 e -> 1 e (not 1.5 e). |
| T6-06 | R 76-1 Table 6 | mpe {"d":1,"e":1,"max":3000,"cls":"III","load":500} | 0.5 | m = 500 e. Class III: 0 <= m <= 500 -> 0.5 e. |
| T6-07 | R 76-1 Table 6 | mpe {"d":1,"e":1,"max":3000,"cls":"III","load":501} | 1 | m = 501 e. Class III: 500 < m <= 2000 -> 1 e. |
| T6-08 | R 76-1 Table 6 | mpe {"d":1,"e":1,"max":3000,"cls":"III","load":2001} | 1.5 | m = 2001 e. Class III: 2000 < m <= 10000 -> 1.5 e. |
| T6-09 | R 76-1 Table 6 | mpe {"d":0.001,"e":0.001,"max":100,"cls":"I","load":50} | 0.5 | m = 50000 e. Class I: 0 <= m <= 50000 -> 0.5 e. |
| T6-10 | R 76-1 Table 6 | mpe {"d":0.001,"e":0.001,"max":400,"cls":"I","load":200.001} | 1.5 | m = 200001 e. Class I: 200000 < m -> 1.5 e. |
| T6-11 | R 76-1 Table 6 | mpe {"d":5,"e":5,"max":3000,"cls":"IIII","load":255} | 1 | m = 51 e. Class IIII: 50 < m <= 200 -> 1 e. |
| T6-12 | R 76-1 Table 6 | mpe {"d":5,"e":5,"max":3000,"cls":"IIII","load":1005} | 1.5 | m = 201 e. Class IIII: 200 < m <= 1000 -> 1.5 e. |
| T6-13 | R 76-1 Table 6 | mpe {"d":0.1,"e":1,"max":3000,"cls":"III","load":1000} | 10 | e = 1 g, d = 0.1 g. m = 1000 e. Class III: 500 < m <= 2000 -> 1 e = 1 g = 10 d. |
| LV-01 | R 76-1 3.5.1, 3.5.2 | level {"err":0.4,"mpe":1,"u":0.5} | "ok" | Within the mpe even in the worst case: 0.4 + 0.5 = 0.9 <= 1. |
| LV-02 | R 76-1 3.5.1, 3.5.2 | level {"err":0.5,"mpe":1,"u":0.5} | "ok" | 0.5 + 0.5 = 1 <= 1 (limit included). |
| LV-03 | R 76-1 3.5.1, 3.5.2 | level {"err":1.2,"mpe":1,"u":0.5} | "maybe" | 1.2 - 0.5 = 0.7 <= 1 < 1.2 + 0.5: the uncertainty does not allow a decision. |
| LV-04 | R 76-1 3.5.1, 3.5.2 | level {"err":1.6,"mpe":1,"u":0.1} | "warn" | Surely above the mpe (1.6 - 0.1 > 1) and surely within the mpe in service = 2 mpe (3.5.2): 1.6 + 0.1 <= 2. |
| LV-05 | R 76-1 3.5.1, 3.5.2 | level {"err":2,"mpe":1,"u":0.1} | "bad" | 2.0 + 0.1 > 2 mpe: not surely within the mpe in service. The in-service tier is only given when it is sure (guard band of u). |
| LV-06 | R 76-1 3.5.1 | level {"err":0,"mpe":0.5,"u":0.5} | "ok" | A rounded reading (u = d/2) with mpe 0.5 d: only an error of 0 is surely within the mpe: 0 + 0.5 <= 0.5. |
| A44-01 | R 76-1 A.4.4.1, Table 3 | loads {"d":0.1,"e":0.1,"max":3000,"cls":"II","purpose":"verify"} | [5,300,500,2000,3000] | At least 5 loads incl. Min (5 g, T3-12, >= 100 mg), the mpe changes 5000 e = 500 g and 20000 e = 2000 g, and Max 3000 g; the fifth is filled in at 10 % of Max = 300 g. |
| A44-02 | R 76-1 A.4.4.1, Table 3 | loads {"d":0.1,"e":0.1,"max":3000,"cls":"II","purpose":"type"} | [5,300,500,600,750,1050,1500,2000,2250,3000] | 10 loads for the initial intrinsic error: Min 5 g, 500 g, 2000 g, Max 3000 g, then 10, 25, 50, 75, 20 and 35 % of Max (300, 750, 1500, 2250, 600, 1050 g), each far enough from the others. |
| A44-03 | R 76-1 A.4.4.1, Table 3 | loads {"d":1,"e":1,"max":3000,"cls":"III","purpose":"verify"} | [20,300,500,2000,3000] | Min = 20 d = 20 g, the mpe changes 500 e = 500 g and 2000 e = 2000 g, Max 3000 g, plus 10 % of Max = 300 g. |
| A410-01 | R 76-1 A.4.10 | reps {"cls":"II","purpose":"verify"} | 6 | Verification: six weighings on classes I and II. |
| A410-02 | R 76-1 A.4.10 | reps {"cls":"III","purpose":"verify"} | 3 | Verification: three weighings on classes III and IIII. |
| A410-03 | R 76-1 A.4.10 | reps {"cls":"III","purpose":"type"} | 10 | Type examination, Max < 1000 kg: 10 weighings per series. |
| A410-04 | R 76-1 A.4.10 | reploads {"max":3000,"purpose":"verify"} | [2400] | Verification: one series with about 0.8 Max = 2400 g. |
| A410-05 | R 76-1 A.4.10 | reploads {"max":3000,"purpose":"type"} | [1500,3000] | Type examination: one series with about 50 % (1500 g) and one close to 100 % of Max (3000 g). |
| A443-01 | R 76-1 A.4.4.3 | rowerr {"d":5,"e":5,"net":1000,"target":1000,"dl":1.5,"zdl":2.5} | {"err":1,"errD":0.2,"u":0} | The example of the standard: e = 5 g, L = 1000 g, I = 1000 g, dL = 1.5 g: E = I + e/2 - dL - L = 1000 + 2.5 - 1.5 - 1000 = 1 g = 0.2 d. The zero, determined the same way, shows 0 with dL0 = 2.5 g: E0 = 0 + 2.5 - 2.5 = 0, so Ec = E - E0 = 1 g; both changeover points remove the rounding (u = 0). (Corrected 2026-10-23: the input now names the zero changeover dL0; the expected values are unchanged.) |
| A443-05 | R 76-1 A.4.4.3 | rowerr {"d":5,"e":5,"net":1000,"target":1000,"dl":1.5,"zdl":2} | {"err":0.5,"errD":0.1,"u":0} | As A443-01, but the zero needs dL0 = 2 g: E0 = 0 + 2.5 - 2 = 0.5 g. Ec = E - E0 = 1 - 0.5 = 0.5 g = 0.1 d. |
| A443-06 | R 76-1 A.4.4.3, 3.5.3.2 | rowerr {"d":5,"e":5,"net":1000,"target":1000,"dl":1.5} | {"err":1,"errD":0.2,"u":0.5} | dL at the load, but no dL0 at the zero: the zero is taken as shown (E0 = 0), its rounding of up to e/2 = 2.5 g stays in the uncertainty: u = 0.5 d. |
| A443-02 | R 76-1 3.5.3.2 | rowerr {"d":0.1,"e":0.1,"net":1000.1,"target":1000,"dl":0} | {"err":0.1,"errD":1,"u":0.5} | A rounded reading without changeover: E = 1000.1 - 1000 = 0.1 g = 1 d, uncertain by half a step: u = 0.5 d. |
| A443-03 | R 76-1 A.4.1.6 | rowerr {"d":0.1,"e":0.1,"net":1000.03,"target":1000,"dl":0,"fine":true} | {"err":0.03,"errD":0.3,"u":0.1} | A reading before rounding (an interval of at most 1/5 e): E = 0.03 g = 0.3 d; u = d/10 = 0.1 d. |
| A443-04 | R 76-1 3.7.1 | rowerr {"d":0.1,"e":0.1,"net":1000.1,"target":1000,"dl":0,"tolPct":0.01} | {"err":0.1,"errD":1,"u":1.5} | A test weight with a tolerance of 0.01 %: 1000 g x 0.0001 = 0.1 g; u = d/2 + 0.1 g = 0.15 g = 1.5 d. |
| W-01 | R 76-1 A.4.4.1, 3.5.1 | weigh {"d":1,"e":1,"max":3000,"cls":"III","fine":true,"up":[[500,500.2],[2000,2000.6],[3000,3001.2]],"down":[[2000,2000.9],[500,500.3]]} | {"checks":[[0.2,0.1],[0.6,0.1],[1.2,0.1],[0.9,0.1],[0.3,0.1],[0.3,0.2],[0.1,0.2]],"status":"meets"} | Readings before rounding (u = 0.1 d each). Errors up 0.2 / 0.6 / 1.2 d against mpe 0.5 / 1 / 1.5 d (500 / 2000 / 3000 e), down 0.9 d at 2000 g (0.9 + 0.1 <= 1) and 0.3 d at 500 g; same load up and down: |0.9 - 0.6| = 0.3 d (u 0.2) and |0.3 - 0.2| = 0.1 d (u 0.2). All surely within: meets class III. |
| R-01 | R 76-1 3.6.1, A.4.10 | repeat {"d":0.1,"e":0.1,"max":3000,"cls":"II","load":2400,"nets":[2400,2400.1,2399.9,2400,2400.1,2400]} | {"checks":[[1,0.5],[2,1]],"status":"maybe"} | Rounded readings (u = 0.5 d). mpe at 2400 g (24000 e, class II) = 1.5 d. Largest error 1 d: 1 + 0.5 <= 1.5 -> within. Difference of the results (3.6.1) = 1 - (-1) = 2 d with u = 1 d: 2 - 1 <= 1.5 < 2 + 1 -> cannot be decided. |
| R-02 | R 76-1 3.6.1, A.4.10 | repeat {"d":0.1,"e":0.1,"max":3000,"cls":"II","load":2400,"fine":true,"nets":[2400.02,2400.05,2399.98,2400.03,2400,2400.04]} | {"checks":[[0.5,0.1],[0.7,0.2]],"status":"meets"} | Readings before rounding (u = 0.1 d). Errors 0.2, 0.5, -0.2, 0.3, 0, 0.4 d: largest 0.5 d <= 1.5 - 0.1. Difference 0.5 - (-0.2) = 0.7 d <= 1.5 - 0.2. |
| R-03 | R 76-1 3.6.1, 3.5.2 | repeat {"d":0.1,"e":0.1,"max":3000,"cls":"II","load":2400,"fine":true,"nets":[2400,2400.2,2399.9,2400,2400.1,2400]} | {"checks":[[2,0.1],[3,0.2]],"status":"fails"} | Largest error 2 d: above 1.5 d, within 2 x 1.5 = 3 d in service (warn). Difference 2 - (-1) = 3 d with u 0.2: 3 > 3 - 0.2 -> not surely within the in-service limit: fails. |
| C-01 | R 76-1 3.9.4.1, 3.9.4.2, A.4.11 | creep {"d":0.1,"e":0.1,"max":3000,"cls":"II","load":2900,"samples":[[0,2900],[300,2900.02],[900,2900.03],[1800,2900.04]],"zeros":[0,0.02]} | {"checks":[[0.4,0.5],[0.1,0.2],[0.2,0.5]],"status":"meets"} | Readings before rounding. Change in 30 min: 0.04 g = 0.4 d <= 0.5 e. Change 15 to 30 min: 0.01 g = 0.1 d <= 0.2 e. Zero return: 0.02 g = 0.2 d <= 0.5 e. |
| C-02 | R 76-1 3.9.4.1 | creep {"d":0.1,"e":0.1,"max":3000,"cls":"II","load":2900,"samples":[[0,2900],[900,2900.01],[1800,2900.04]],"zeros":[0,0.01]} | {"checks":[[0.4,0.5],[0.3,0.2],[0.1,0.5]],"status":"fails"} | 30 min: 0.4 d <= 0.5 e, but 15 to 30 min: 0.03 g = 0.3 d > 0.2 e -> the 30 minute test is not met (the 4 hour alternative would be next). |
| Z-01 | R 76-1 4.5.2, A.4.2.3.1 | zset {"d":0.1,"e":0.1,"max":3000,"cls":"II","fine":true,"nets":[0.01]} | {"checks":[[0.1,0.25]],"status":"meets"} | After the zero-setting the indication is 0.01 g = 0.1 d, u = 0.1 d: 0.1 + 0.1 <= 0.25 e. |
| Z-02 | R 76-1 4.5.2 | zset {"d":0.1,"e":0.1,"max":3000,"cls":"II","fine":true,"nets":[0.03]} | {"checks":[[0.3,0.25]],"status":"maybe"} | 0.3 d with u = 0.1 d against 0.25 e: 0.3 - 0.1 <= 0.25 < 0.3 + 0.1 -> cannot be decided. |
| Z-03 | R 76-1 4.5.2 | zset {"d":0.1,"e":0.1,"max":3000,"cls":"II","fine":true,"nets":[0.05]} | {"checks":[[0.5,0.25]],"status":"fails"} | 0.5 d - 0.1 > 0.25 e: the zero-setting is not accurate enough (no in-service tier for this limit). |
| S-01 | R 76-1 4.4.2, A.4.12 | stab {"d":0.1,"e":0.1,"max":3000,"cls":"II","spans":[0.1,0.2]} | {"checks":[[1,1],[2,1]],"status":"fails"} | After the stable signal the values may differ by at most 1 e (two adjacent values): a span of 0.1 g = 1 d is within, 0.2 g = 2 d is not. |
| D-01 | R 76-1 3.8.2.2, A.4.8.2 | disc {"d":0.1,"e":0.1,"max":3000,"cls":"II","rows":[[1500,1500,1500.1],[5,5,5]]} | {"checks":[[0,0.5],[1,0.5]],"status":"fails"} | After the extra 1.4 d the indication must be I + d: at 1500 g it shows 1500.1 = I + d (0 d off); at 5 g it still shows 5.0 = I (1 d off): the discrimination is not met. |
| LIN-01 | least squares (EURAMET cg-18 style, R 76 has no linearity limit) | linearity {"pts":[[500,500.05],[1000,1000.1],[2000,2000.2],[3000,3000.3]]} | {"a":1.0001,"maxDev":0} | All points lie on y = 1.0001 x: slope 1.0001 (gain error +0.01 %), no deviation from the line. |
| LIN-02 | least squares | linearity {"pts":[[1,1],[2,2],[3,4]]} | {"a":1.5,"maxDev":0.333333333} | n = 3, sum x = 6, sum y = 7, sum xy = 17, sum x2 = 14: a = (3 x 17 - 6 x 7) / (3 x 14 - 36) = 9 / 6 = 1.5, b = (7 - 1.5 x 6) / 3 = -2/3; the line gives 0.833, 2.333, 3.833: largest deviation 1/3 at x = 2. |
| ECC-01 | R 76-1 3.6.2, Table 6 (difference to the middle, EURAMET cg-18) | ecc {"d":0.1,"e":0.1,"max":3000,"cls":"II","fine":true,"rounds":[[1000,1000.05,999.97,1000.07,1000.01]]} | {"maxD":0.7,"mpe":1,"u":0.2,"level":"ok"} | 3 rounds of the same values: corner means differ from the middle by 0.5, -0.3, 0.7, 0.1 d; largest 0.7 d. mpe at 1000 g (10000 e, class II) = 1 e. Two means before rounding: u = 0.2 d. 0.7 + 0.2 <= 1. |
| ECC-02 | R 76-1 3.6.2, Table 6 | ecc {"d":0.1,"e":0.1,"max":3000,"cls":"II","fine":false,"rounds":[[1000,1000.1,1000,999.9,1000]]} | {"maxD":1,"mpe":1,"u":1,"level":"maybe"} | Rounded readings: the difference of two rounded means is uncertain by up to 1 d. Largest difference 1 d: 1 - 1 <= 1 < 1 + 1 -> cannot be decided. |
| ECC-03 | R 76-1 3.6.2, 3.5.2 | ecc {"d":0.1,"e":0.1,"max":3000,"cls":"II","fine":true,"rounds":[[1000,1000.25,1000,1000,1000]]} | {"maxD":2.5,"mpe":1,"u":0.2,"level":"bad"} | 2.5 d + 0.2 > 2 x 1 d: above even the in-service limit. |
| ECC-04 | R 76-1 3.6.2, A.4.7 (against the true mass) | ecc {"d":0.1,"e":0.1,"max":3000,"cls":"II","fine":true,"mass":999.98,"rounds":[[1000,1000.03,999.95,1000.02,1000]]} | {"absD":0.5,"absU":0.1,"absMpe":1,"absLevel":"ok"} | True mass 999.98 g. Corner means 1000.03, 999.95, 1000.02, 1000.00 g: errors +0.05, -0.03, +0.04, +0.02 g; largest 0.05 g = 0.5 d (the middle is not judged, A.4.7). Readings before rounding: u = d/10 = 0.1 d. mpe at 999.98 g (9999.8 e, class II) = 1 e. 0.5 + 0.1 <= 1. |
| ECC-05 | R 76-1 3.6.2, A.4.7, 3.7.1 | ecc {"d":0.1,"e":0.1,"max":3000,"cls":"II","fine":false,"mass":1000,"tolPct":0.005,"rounds":[[1000,1000.1,1000,1000.1,1000]]} | {"absD":1,"absU":1,"absMpe":1,"absLevel":"maybe"} | Rounded readings: u = d/2 = 0.5 d, plus the tolerance of the weight 1000 g x 0.00005 = 0.05 g = 0.5 d: u = 1 d. Largest corner error 0.1 g = 1 d against mpe 1 e: 1 - 1 <= 1 < 1 + 1 -> cannot be decided. |
| ECC-06 | R 76-1 3.6.2, 3.5.2 | ecc {"d":0.1,"e":0.1,"max":3000,"cls":"II","fine":true,"mass":1000,"rounds":[[1000,1000.12,1000,1000,1000]]} | {"absD":1.2,"absU":0.1,"absMpe":1,"absLevel":"warn"} | Largest corner error 0.12 g = 1.2 d: surely above 1 d (1.2 - 0.1 > 1), surely within the in-service limit 2 d (1.2 + 0.1 <= 2). |
| ECC-07 | R 76-1 A.4.7 | ecc {"d":0.1,"e":0.1,"max":3000,"cls":"II","fine":true,"mass":1000,"rounds":[[1000.3,1000,1000,1000,1000]]} | {"absD":0,"absLevel":"ok"} | Only the middle is off (0.3 g); A.4.7 loads only the eccentric segments, so the R 76 result looks at the corners alone: 0 d. (The difference to the middle, 3 d, is still shown for information.) |
| W111-01 | R 76-1 3.7.1 | weighttol {"d":0.1,"e":0.1,"max":3000,"cls":"II","load":2000,"tolPct":0.005} | false | mpe at 2000 g (20000 e, class II) = 1 e = 0.1 g; 1/3 of it = 0.033 g. The weight may be off by 2000 x 0.00005 = 0.1 g > 0.033 g: not allowed. |
| W111-02 | R 76-1 3.7.1 | weighttol {"d":0.1,"e":0.1,"max":3000,"cls":"II","load":2000,"tolPct":0.0015} | true | 2000 x 0.000015 = 0.03 g <= 0.033 g: allowed. |
| W111-03 | R 76-1 3.7.1 | weighttol {"d":0.1,"e":0.1,"max":3000,"cls":"II","load":2000,"tolPct":0.002} | false | 2000 x 0.00002 = 0.04 g > 0.033 g (1/3 of the mpe): not allowed, although it is below half of the mpe. |
| RT-01 | USP <41> repeatability + accuracy | routine {"d":0.1,"e":0.1,"max":3000,"p":{"crit":"lab","w1":100,"w2":0,"mSmall":100,"repPct":0.1,"accPct":0.1},"v1":[100,100.1,100,100,100.1,100,100,100.1,100,100],"v2":[]} | {"mean":100.03,"s":0.048304589,"minW":96.609178,"checks":[[0.966091783,1],[0,1]],"pass":true} | Mean 100.03 g. Deviations +0.07 g (3 x) and -0.03 g (7 x): sum of squares 3 x 0.0049 + 7 x 0.0009 = 0.021; s = sqrt(0.021 / 9) = 0.0483046 g (above 0.41 d = 0.041 g, so it is used). 2 s = 0.0966 g = 0.966 d <= 0.10 % of the smallest net weight 100 g = 0.1 g = 1 d. Minimum weight 2000 s = 96.61 g. Accuracy: the first weighing 100.0 g (no second weight) - error 0 d <= 0.10 % of 100 g = 1 d. PASS. |
| RT-02 | USP <41> repeatability (s replaced by 0.41 d) | routine {"d":0.1,"e":0.1,"max":3000,"p":{"crit":"lab","w1":50,"w2":0,"mSmall":50,"repPct":0.1,"accPct":0.1},"v1":[50,50,50,50,50,50,50,50,50,50],"v2":[]} | {"s":0,"sEff":0.041,"minW":82,"checks":[[0.82,0.5],[0,0.5]],"pass":false} | All ten readings equal: s = 0, replaced by 0.41 d = 0.041 g. 2 x 0.041 = 0.082 g = 0.82 d > 0.10 % of 50 g = 0.05 g = 0.5 d: FAIL - the smallest net weight is below the minimum weight 2000 x 0.041 = 82 g. |
| RT-03 | Ph. Eur. 2.1.7 sensitivity | routine {"d":0.1,"e":0.1,"max":3000,"p":{"crit":"lab","w1":100,"w2":1000,"mSmall":200,"repPct":0.1,"accPct":0.05},"v1":[100,100,100,100,100,100,100,100,100,100],"v2":[1000.6]} | {"checks":[[0.82,2],[6,5]],"pass":false} | Repeatability: s = 0 -> 0.041 g; 2 s = 0.082 g = 0.82 d <= 0.10 % of 200 g = 0.2 g = 2 d. Sensitivity: 1000.6 - 1000 = 0.6 g = 6 d > 0.05 % of 1000 g = 0.5 g = 5 d: FAIL. |
| RT-04 | Routine check, own tolerance (quick profile) | routine {"d":1,"e":1,"max":3000,"p":{"crit":"tol","w1":1000,"w2":0,"errPct":1,"errMinD":2},"v1":[1004,1006,1005],"v2":[]} | {"mean":1005,"checks":[[5,10]],"pass":true} | Mean 1005 g, error 5 g = 5 d. Tolerance: the larger of 1 % of 1000 g = 10 g and 2 d = 2 g: 10 g = 10 d. 5 <= 10: PASS. |
| RT-05 | Routine check, own tolerance (minimum 2 d) | routine {"d":1,"e":1,"max":3000,"p":{"crit":"tol","w1":100,"w2":0,"errPct":1,"errMinD":2},"v1":[103,103,102],"v2":[]} | {"mean":102.666667,"checks":[[2.666667,2]],"pass":false} | Mean 308 / 3 = 102.667 g, error 2.667 d. Tolerance: the larger of 1 % of 100 g = 1 g and 2 d = 2 g: 2 d. 2.667 > 2: FAIL. |
| RT-06 | R 76-1 3.5.2 (in use = 2 x mpe of Table 6) | routine {"d":1,"e":1,"max":3000,"cls":"III","p":{"crit":"mpe2","w1":1000,"w2":2500},"v1":[1001,1002,1000],"v2":[2503]} | {"checks":[[2,2],[3,3]],"pass":true} | Class III, e = 1 g: 1000 g = 1000 e (500 < m <= 2000): mpe 1 e, in use 2 e; worst value 1002 g = 2 d <= 2. 2500 g = 2500 e (> 2000): mpe 1.5 e, in use 3 e; 2503 g = 3 d <= 3. PASS (limits included). |
| DS-01 | R 76-1 Table 3; NIST HB 44 S.1.2.1 (d = 1, 2 or 5 x 10^n) | classes {"d":0.2,"e":0.2,"max":2000} | ["II","III"] | d = 0.2 g (a 2 x 10^-1 step). n = 2000 / 0.2 = 10000. I: n < 50000. II with e >= 0.1 g needs 5000 <= n <= 100000: yes. III with 0.1 <= e <= 2 g needs 100 <= n <= 10000: yes (limit included). IIII needs e >= 5 g: no. |
| DS-02 | R 76-1 Table 3, 3.4.3 | min {"d":0.2,"e":0.2,"max":2000,"cls":"II"} | 10 | Class II with e >= 0.1 g: Min = 50 d = 50 x 0.2 g = 10 g. |
| DS-03 | R 76-1 Table 3 | classes {"d":20000,"e":20000,"max":60000000} | ["III"] | A weighbridge: Max 60 t, d = e = 20 kg. n = 60000000 / 20000 = 3000. II with e >= 0.1 g needs n >= 5000: no. III with e >= 5 g needs 500 <= n <= 10000: yes. IIII needs n <= 1000: no. |
| DS-04 | R 76-1 Table 3 | classes {"d":1.0e-7,"e":1.0e-7,"max":2.1} | [] | An ultra-micro balance: Max 2.1 g, d = e = 0.1 ug = 0.0000001 g, n = 21000000. Class I needs e >= 0.001 g: no; II needs e >= 0.001 g: no; III and IIII need e >= 0.1 g: no. No class of R 76 covers it: the tests are run, the report names no class that fits. |
| DS-05 | R 76-1 Table 3 | classes {"d":1.0e-6,"e":1.0e-6,"max":6.1} | [] | A microbalance: Max 6.1 g, d = e = 1 ug = 0.000001 g, n = 6100000. All classes need e >= 0.001 g: no class fits. |
| MI-01 | R 76-1 3.3 / Table 3 for each partial range (multi-interval; our reading: a class fits when it fits every range) | classes {"d":2,"e":2,"max":15000,"ranges":[{"d":2,"max":6000},{"d":5,"max":15000}]} | ["III"] | Retail scale 6 / 15 kg, d = e = 2 / 5 g. Range 1: n1 = 6000 / 2 = 3000; range 2: n2 = 15000 / 5 = 3000. II needs n >= 5000 (e >= 0.1 g): no. III: range 1 with 0.1 <= e <= 2 g needs 100 <= n <= 10000: yes; range 2 with e >= 5 g needs 500 <= n <= 10000: yes. IIII needs e >= 5 g in range 1 (e1 = 2 g): no. |
| MI-02 | R 76-1 Table 6 (mpe in e of the partial range of the load) | mpe {"d":2,"e":2,"max":15000,"cls":"III","load":10000,"ranges":[{"d":2,"max":6000},{"d":5,"max":15000}]} | 2.5 | 10000 g is above 6000 g + 1 g (Max 1 + d1 / 2): range 2, e = 5 g. 10000 / 5 = 2000 e: 500 < m <= 2000 -> mpe 1 e = 5 g. In d of range 1 (2 g): 5 / 2 = 2.5 d. |
| MI-03 | R 76-1 Table 6 (mpe in e of the partial range of the load) | mpe {"d":2,"e":2,"max":15000,"cls":"III","load":5000,"ranges":[{"d":2,"max":6000},{"d":5,"max":15000}]} | 1.5 | 5000 g is in range 1, e = 2 g. 5000 / 2 = 2500 e: m > 2000 -> mpe 1.5 e = 3 g = 1.5 d (d of range 1 = 2 g). |
| MI-04 | R 76-1 Table 6, the boundary of a partial range | mpe {"d":2,"e":2,"max":15000,"cls":"III","load":6002,"ranges":[{"d":2,"max":6000},{"d":5,"max":15000}]} | 2.5 | 6002 g is above 6000 g + 1 g: range 2 (e = 5 g). 6002 / 5 = 1200.4 e: 500 < m <= 2000 -> mpe 1 e = 5 g = 2.5 d of range 1. (At 6000 g the load is still in range 1: 3000 e -> 1.5 e = 3 g = 1.5 d.) |
| MI-05 | R 76-1 A.4.4 (error of a reading, rounding d / 2 of the range of the load) | rowerr {"d":2,"e":2,"max":15000,"net":10005,"target":10000,"ranges":[{"d":2,"max":6000},{"d":5,"max":15000}]} | {"err":5,"errD":2.5,"u":1.25} | Error 10005 - 10000 = 5 g = 2.5 d of range 1 (2 g). The reading is rounded to d of range 2 (5 g): uncertainty 5 / 2 = 2.5 g = 1.25 d of range 1. |
Also useful for you
QA guide: validation (IQ/OQ/PQ)
Labs, quality & metrologyIntroducing it in a regulated lab: the QA tasks, what the app takes over, IQ / OQ / PQ, an estimate of the effort and the steps to the sign-off.
Read →Scale calibration & tests
Labs, quality & metrologyEccentricity, linearity, repeatability and every other test of OIML R 76-1 Annex A at the scale, with checklist and final report.
Read →Pipette calibration
Labs, quality & metrologyGravimetric pipette calibration along ISO 8655-6, step by step on your lab balance: the scale takes every value, the uncertainty, the CMC, the register.
Read →Daily balance check (USP <41>)
Labs, quality & metrologyUSP <41> repeatability and minimum weight, Ph. Eur. 2.1.7, trade and HACCP checks in one minute a day.
Read →Who else uses it
Sealed records & checks: more pages
Sealed or signed?
Labs, quality & metrologyThe seal of the scale and the signature of a person: what each one says and how anybody checks it.
Read →Tamper-evident records
Labs, quality & metrologyAudit trail, hash chain, numbered records, the LEGAL MODE store - and the controls an audit expects.
Read →For auditors & officials
Labs, quality & metrologyWhat you check yourself without an account or a scale - and what to ask the lab or the shop for.
Read →The same app, for other work
Every function is in the same app - at home, in business and in the lab. All functions at a glance.