UP↑SCALE › LABS, QUALITY & METROLOGY › SERVICE TECHNICIANS & INSPECTORS › SCALE CALIBRATION & TESTS
Scale calibration and test suite - at the scale, on your tablet
A scale test suite is the set of tests that shows whether a scale weighs within the limits of its accuracy class. UP↑SCALE runs every test of OIML R 76-1 Annex A that can be done at the scale, guided step by step, with the limits worked out for the class of your scale - and a report you can sign. For owners who want to know, and for technicians and inspectors who test for others.
- Eccentricity, linearity, repeatability, creep and more
- Limits of OIML R 76 or NIST HB 44 for your class
- A strict procedure along OIML R 76-1 Annex A
- Numbered, sealed test reports and certificates
In short
UP↑SCALE is one scale app for the scale you already have. Its test suite runs the tests of OIML R 76-1 Annex A at the scale - eccentricity, linearity, repeatability and more - against the limits of OIML R 76 or NIST HB 44, for scale owners, labs and service technicians. The simple tests are free, the extended ones PRO; sealed reports and cg-18 certificates need ENTERPRISE (144.00 EUR a year).
What it saves you
The tests: eccentricity, linearity, repeatability, creep and more FREEPRO
| Test | What it checks | Limit (in short) |
|---|---|---|
| Repeatability R 76-1 A.4.10 | the same load several times - the spread (6 / 3 weighings by class for a verification, 10 for a type examination) | mpe of the class |
| Accuracy & linearity A.4.4 | weighing up and down with the loads of the standard (Min, the mpe change points, Max); best-fit line deviation | mpe at each load |
| Eccentricity | the same load in the corners and the middle | mpe at the load |
| Tare | weighing after a tare | mpe of the net load |
| Zero-setting A.4.2.3.1 | load on, zero-set, the reading at zero; and the zero indicator | ± 0.25 e |
| Discrimination A.4.8.2 | at Min, ½ Max and Max: adding 1.4 d must change the reading (type examination, d ≥ 5 mg) | the reading follows |
| Creep & zero return A.4.11 | a load for 30 minutes (or the 4 h alternative), then zero | ≤ 0.5 e (30 min), ≤ 0.2 e (15-30 min), zero return ≤ 0.5 e |
| Warm-up A.5.2 | zero and load at 0, 5, 15 and 30 minutes after switching on (after 8 hours off) | mpe |
| Stable equilibrium A.4.12 | a disturbance, then the first stable signal: the values of 5 s | spread ≤ 1 e |
| Calibration check | known weights against the reading | your limits |
Free: the routine check, the calibration check, repeatability, eccentricity and the weighing test with the default loads, results on the screen. PRO: tare, zero-setting, discrimination, creep, warm-up, stable equilibrium and the strict procedure, and a plain print of a single test. A scale without a link? Type its readings: the UP↑SCALE or the free virtual scale works them out. Each test runs with the filters it needs: raw values, or the legal display "as used" where the standard asks for it.
Manual: the tests, filters per test, eccentricity, calibration check
The tests in the app
Every load up and down with its error, the limit and the check - and the corner load test, judged against OIML R 76.


More about the app, with all screenshots →Screenshots of the real app with demo values - a screenshot may show the look of an older app version.
Maximum permissible errors and accuracy classes - worked out for you
- the accuracy classes (I, II, III, IIII) that fit d, e and Max of your scale - also for every range of a multi-interval scale
- the maximum permissible errors at each load, for verification and in use
- the uncertainty of the readings (the rounding to d), the test-weight tolerance (R 111 weights, at most 1/3 of the mpe: R 76-1 3.7.1), the room temperature change and the warm-up time
- every rule shown as ✓ met, ✗ not met or ? not known - nothing hidden
- the maths is checkable: the reference cases of the validation are public
A scale test and calibration checklist per procedure
The overview shows what is done, with which result, and which test is asked for:
- Verification / conformity assessment - the tests of R 76-1 Annex A
- Inspection in use - a typical set (the national rules decide)
- Calibration - along EURAMET cg-18: repeatability, eccentricity, errors of indication
Manual: checklist
The strict procedure: OIML R 76-1 Annex A as closely as a self-check can PRO
One switch, and the suite follows Annex A as closely as a self-check can: the loads of the standard, a preload to Max, the legal display with the ΔL input at every row (changeover point, A.4.4.3), repeatability with the number of weighings of the class, the warm-up time from the uptime of the scale, the tolerance of your test weights and the temperatures - with its own checklist in the report.
Manual: strict procedure
OIML R 76 / EN 45501 or NIST HB 44 - verification or in service FREE
Pick the region - EU / EEA, Germany, Switzerland, United Kingdom or USA - and what the test is for: verification (the mpe of a first verification, HB 44 acceptance tolerance) or in-service inspection (twice the mpe, HB 44 maintenance tolerance). The UP↑SCALE judges by that standard. Where both standards give the same result, the report says "OIML R 76 / NIST HB 44"; where HB 44 differs (the fourth tolerance step of classes III and IIII, the repeatability and shift ranges, discrimination, zero return), it says which one it applied. Not covered: class III L, wheel-load, axle-load, crane and hopper scales.
Manual: standard and purpose, typed readings
Scale calibration along EURAMET cg-18, with certificates ENTERPRISE
- Calibration with uncertainty: repeatability, eccentricity, the error of indication at every load with its expanded uncertainty U (k from the effective degrees of freedom), substitution loads where the weights do not reach, and the uncertainty in use when the fit passes - all computed on the UP↑SCALE
- Any maker's scale: readings from the connected scale or typed in; the register of the customers' scales with due dates and a due list
- Test weights register: OIML R 111, ASTM E617 and NIST class F weights with certificate, issuer, correction and U per piece
- Certificates of your lab: standard, or with your own accreditation (your symbol image, your CMC, results outside marked *); the app checks completeness against the listed rules; procedures, signatories with approval, a certificate register with amendments and withdrawal - pipette certificates the same way
- Service documents: the Placed in Service Report (US), the repair and test protocol for the verification authority (DE), the verification certificate of a private body (AT, ES, FR, IT, BE, SE) - the document of the agency, company or body that issues it
- One data file moves the registers, runs and certificates to another browser, part by part: merge, or a clean overwrite that keeps every history
Manual: scale calibration, certificates, service documents, data file
Test and calibration reports PROENTERPRISE
- Plain print of a single test with PRO; with ENTERPRISE the test report per test - print, PDF, HTML, CSV with all readings
- Final report on certificate-style pages: scale data and settings, checklist, summary per procedure, all tables with the single readings, the class overview, date, signature, stamp and remarks
- Your data on every printout: name, organisation, inspector ID, place, report number, remarks and a signature drawn once on the screen
- Numbered and sealed by the scale - checkable on up-scales.com/check
- Test history and register on the scale: interval, due date, trend
Compared with test weights and a paper form
| Test weights + paper form | UP↑SCALE | |
|---|---|---|
| Readings | read and written by hand | taken from the scale when stable |
| Limits | looked up in tables | worked out for the class, shown per rule |
| Forgotten tests | possible | the checklist shows what is open |
| The report | typed later | ready at the scale, numbered and sealed |
What it saves in hours and subscriptions for a whole lab: the cost table of the QA guide.
What it is not
- A test record the signing person is responsible for - no official verification, no logos of an authority. The calibration certificates and service documents (ENTERPRISE) are the documents of your lab, company or body: the app fills them in and checks their completeness, the content stays your responsibility
- It does not replace your reference weights, your procedure or your decision
- With the RS-232 adapter on a verified scale the display of that scale stays the legal reading; the adapter is no part of the legal measuring chain
- Whether a record is accepted is the decision of the responsible authority
Scale calibration and testing: questions and answers
How do I do an eccentricity test (corner load test) on a balance?
Put the same load in the middle and then in each corner of the platform and compare the readings. Each one must stay within the maximum permissible error at that load. The app guides the positions, takes the values when stable and shows each error in d against the OIML R 76 limit.
How many weighings does a repeatability test need?
Along OIML R 76-1 A.4.10: 6 or 3 weighings, depending on the class, for a verification, and 10 for a type examination. The spread must stay within the mpe of the class; the strict procedure sets the number for you.
What is the maximum permissible error of a weighing scale?
It depends on the accuracy class and the load, in steps of e: for class III 0.5 e up to 500 e, 1 e up to 2000 e and 1.5 e above - for verification; the limit in use is twice that. The class (I, II, III or IIII) follows from e, d and Max of the scale. The app works out the mpe at every load, for verification and in use - also for each range of a multi-interval scale - and shows every rule as met, not met or not known.
How do I do a linearity test on a balance?
Weigh loads step by step up to Max and down again, then the empty plate: the error at every load must stay within the maximum permissible error, going up and going down. OIML R 76-1 has no separate linearity test - it is the weighing test of A.4.4, with at least 5 loads (10 for a type examination) including Min, the loads where the mpe changes and Max. The app also shows the largest deviation from the best-fit straight line.
How do I check the accuracy of a scale with a test weight?
Put a known test weight on the middle of the plate 3 times (classes III and IIII) or 6 times (classes I and II): the calibration check takes each value minus the empty plate before it and gives the mean difference in the unit, in d and in %. Up to 0.05 % of the weight (at least 2 d) the calibration is right, up to 0.1 % (at least 5 d) acceptable - and the mean error is also judged against the limits of the classes.
Which tests does a balance calibration include?
Along EURAMET cg-18: repeatability, eccentricity and the errors of indication. The checklist of the test suite marks them for a calibration, next to the tests of OIML R 76-1 Annex A for a verification or conformity assessment and a typical set for an inspection in use (the national rules decide), and shows which of the asked tests are still open.
How long does a creep test take?
A load stays on for 30 minutes (or the 4-hour alternative): the change must be at most 0.5 e in 30 min and at most 0.2 e between 15 and 30 min, and the zero return at most 0.5 e.
Is the test report a certificate?
The test report itself is a test record the signing person is responsible for - no official verification. With ENTERPRISE & COMPLIANCE the app also issues the calibration certificate of your lab along EURAMET cg-18 (standard, or accredited if your lab holds its own accreditation) and fills in service documents: the Placed in Service Report of a US registered service agency, the repair and test protocol for the German verification authority, the verification certificate of a private body in Austria, Spain, France, Italy, Belgium or Sweden. Each is the document of the issuer - the app checks its completeness against the listed rules, numbers and seals it. With the RS-232 adapter on a verified scale the display of that scale stays the legal reading. For service technicians.
What does the scale test suite cost?
The simple tests (routine, calibration check, repeatability, eccentricity, weighing test) and the choice of OIML R 76 or NIST HB 44 are free. PRO adds the extended tests and a plain print of a single test. The numbered reports - test reports, the final report on certificate-style pages, numbered and sealed by the scale - the cg-18 calibration and the certificates - need ENTERPRISE & COMPLIANCE (144.00 EUR a year or 599.00 EUR once).
Can I test a used scale before I buy it?
Yes. Plug in the adapter, run repeatability, eccentricity and linearity on the spot with a few test weights, and you know whether the scale holds its class - before the money is gone.
Who else uses it
Service technicians & inspectors
Labs, quality & metrologyThe tests of OIML R 76-1 Annex A or NIST HB 44 at the scale, EURAMET cg-18 calibrations with the certificate of your lab, service documents after a repair - recorded while you test, numbered and sealed.
Read →Business & production
Business & productionWeighing scale software for shops, packers, kitchens and workshops: weigh, price, label, count, check and book on the scale you have - pay once.
Read →Food producers & packers
Business & productionFill, check and label your packs on the scale you already have - with the statistics and the ℮ lot report the inspector wants to see.
Read →Manufacturers & workshops
Business & productionCount parts on several scales, sort by weight, fill molds and drive your filling machine with a cut-off that learns.
Read →Owners of verified trade scales
Business & productionKeep your legal-for-trade scale exactly as it is - and add everything it never had. The adapter shows the exact weight of your scale and sends only what its keys do (tare, print) - or nothing at all.
Read →Jewellers & gold buyers
Business & productionDensity measurement guided by the scale to spot fakes, a gold calculator with karat and spot price, and every gram on record.
Read →Laboratories & pharmacies
Labs, quality & metrologyThe compounding assistant from master formula to signed batch record, routine checks before use, a full test suite and records you can prove: users, roles, reasons, a hash-chained audit trail and sealed reports.
Read →Auditors & QA teams
Labs, quality & metrologyCheck the records of many scales in minutes instead of days: every file checks itself, the reviews and signatures are in the records, and the documents for the binder come out of the app.
Read →Daily balance check (USP <41>)
Labs, quality & metrologyUSP <41> repeatability and minimum weight, Ph. Eur. 2.1.7, trade and HACCP checks in one minute a day.
Read →Calculation validation (IQ/OQ)
Labs, quality & metrologyReference cases worked out by hand from the standards, computed by the scale firmware, with a sealed record.
Read →The same app, for other work
Every function is in the same app - at home, in business and in the lab. All functions at a glance.